New TopShim run:

version = topshim 2.0.0
date = 2023-11-21
time = 16:26:37.550
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.6

initial B0 stdDev = 2.79 Hz
sample size = 2.92 cm, position = -0.02 cm
final B0 stdDev = 0.32 Hz > improvement = 8.6
envelope width = 0.72 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+65
Z2	-38
Z3	+113
Z4	-213
Z5	+42
X	-27
Y	+130

duration = 3 min 56 sec
completed successfully
finished Tue Nov 21 16:30:32 2023


========================================================================